Particle size is a critical quality attribute for a diverse array of pharmaceutical products, from topical ointments to powders
for pulmonary delivery. During recent decades, the unique attributes of laser-diffraction analysis have positioned it as the
particle-sizing technique of choice for the resulting spectrum of pharmaceutical applications. Fast, nondestructive and suitable
for a broad size range (0.1 to 3000 μm), laser diffraction lends itself to full automation. As a result, for many routine
users, particle-size measurement is now simply a matter of loading the sample and pressing a button. However, streamlining
measurement to this degree demands the development of a robust secure method that will consistently deliver reliable and reproducible
data. This process is given considerable emphasis in the International Organization for Standardization's (ISO) 13320:2009
standard for laser diffraction released at the end of 2009, reflecting that understanding of this technique among industry
has grown significantly during the past decade (1). There is a wealth of information to support method development, the success
of which depends on a rigorous and systematic examination of the factors known to influence results.
The principles of laser diffraction
Understanding the basic principles of laser diffraction is essential for successful method development. Laser diffraction
is an ensemble particle-sizing technique, which means it provides a result for the whole sample, rather than building up distributions
from data for individual particles, in the way that, for example, image analysis or microscopy does. Particles illuminated
in a collimated laser-beam scatter light over a range of angles. Large particles generate a high scattering intensity at relatively
narrow angles to the incident beam, while smaller particles produce a lower intensity signal but at much wider angles. Using
an array of detectors, laser-diffraction analyzers record the pattern of scattered light produced by the sample.
With the application of an appropriate model of light behavior the particle-size distribution of the sample can be determined
from the scattering data, via a deconvolution step. The Mie theory and the Fraunhofer approximation (of Mie theory) are used routinely. ISO 13320:2009
provides a detailed description of both models but confirms Mie as the method of choice, especially for measurements across
a wide dynamic range. The two models return similar results for large particle sizes, but Mie offers improved accuracy for
finer materials. Furthermore, the inaccuracies that arise from the use of Fraunhofer are unpredictable. Both models assume
that the measured particles are spherical, so for nonspherical samples the size distribution returned is one that is based
on spherical equivalence.
Method development involves addressing the wider set of practical issues that flow from this underlying explanation of the
basic principles of operation. United States Pharmacopeia (USP) General Chapter <429> states that laser diffraction involves the measurement of "a representative sample, dispersed at an
adequate concentration in a suitable liquid or gas" (2). This statement neatly highlights three crucial aspects of laser-diffraction
analysis: sampling, dispersion, and measurement conditions.
Sampling
Obtaining a representative sample from a larger bulk is a major challenge with any kind of laboratory-based particle characterization
technique. Sampling issues generate the greatest errors in laser-diffraction analysis, especially when measuring large particles
or when the specification is based on size parameters close to the extremes of the distribution, such as the Dv95 (the particle
size below which 95% of the volume of particles exists). This is because laser diffraction, as a volume-based technique, is
extremely sensitive to small changes in the amount of coarse particles within the selected sample. ISO 13320:2009 expressly
discourages the use of specifications based on the Dv100 for this reason. The effect of sampling on reproducibility increases
with particle size and the width of the distribution, as the volume of sample required to ensure representative sampling of
the coarse particle fraction increases. For this reason, it may be necessary to measure a large sample (often greater than
1–2 g) to ensure reproducible results. For wet-dispersion measurements, this will require the use of a large dispersant volume,
to ensure the concentration of material is within the required range for accurate measurements (i.e., to ensure multiple scattering
is avoided). A detailed discussion of sampling is beyond the remit of this paper, but ISO 14488 provides an extremely useful
summary of the requirements for sampling during particle characterization studies (3).