Increasing the Life-Span of a Probe to Cover a Warranty Period

Equipment and Processing Report

Untitled Document

Q. I work in a quality control laboratory and use a pH meter with a single-junction reference probe to test samples taken from various pharmaceutical (drug) production lines. The probe I use has a six-month warranty. It slowly stops working after just a couple weeks. I am replacing it once a month. We make sure to store it correctly according to the manufacturer’s specifications and even use probe cleaner at the end of each shift to ensure that it stays clean. What can I do to increase the lifespan to at least cover the warranty period?

A. The most common problem seen with pH readings in pharmaceutical-drug testing is a progressively sluggish response. If a probe calibrates quickly but does not respond well in a sample or if it slowly quits working within a few weeks, it is most likely that the probe is not compatible with the sample. Single-junction probes are good for measuring weak ionic water-based solutions and drinking water. Because you are working with drugs, which are much more complex molecules, single-junction references are not suitable. There is nothing that can be done to prolong the life of a single-junction probe in this situation; they are simply not compatible.

Double and flushable junction probes are more appropriate for pharmaceutical applications because samples tend to be more complex. Large molecules will clog the single-junction probes over time. Thus, they may work well initially then become slow to respond over time. If a double junction probe is responding sluggishly, it simply may need to be cleaned. You can continue to clean the probe as you have (i.e., daily). Cleaning prevents proteins, fatty acids, and oils from building up on the glass and causing a delayed response. Proper cleaning and storage of probes is imperative for obtaining the longest life and their optimal accuracy.

If you encounter a situation in which the probe does not change the measurement values at all, the probe has probably shorted out. This happens most often during shipping or with rough handling. The wires used for the cable and reference are thin and somewhat fragile, and will give a steady reading if broken. We have also seen this happen often in production and manufacturing environments.

—Jennifer Minor-Newton, technical application specialist for Cole-Parmer

 

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters

Subscribe: Click to learn more about the newsletter
| Weekly
| Monthly
|Monthly
| Weekly

Survey
FDASIA was signed into law two years ago. Where has the most progress been made in implementation?
Reducing drug shortages
Breakthrough designations
Protecting the supply chain
Expedited reviews of drug submissions
More stakeholder involvement
Reducing drug shortages
32%
Breakthrough designations
11%
Protecting the supply chain
37%
Expedited reviews of drug submissions
11%
More stakeholder involvement
11%
View Results
Jim Miller Outsourcing Outlook Jim Miller Health Systems Raise the Bar on Reimbursing New Drugs
Cynthia Challener, PhD Ingredients Insider Cynthia ChallenerThe Mainstreaming of Continuous Flow API Synthesis
Jill Wechsler Regulatory Watch Jill Wechsler Industry Seeks Clearer Standards for Track and Trace
Siegfried Schmitt Ask the Expert Siegfried SchmittData Integrity
Sandoz Wins Biosimilar Filing Race
NIH Translational Research Partnership Yields Promising Therapy
Clusters set to benefit from improved funding climate but IP rights are even more critical
Supplier Audit Program Marks Progress
FDA, Drug Companies Struggle with Compassionate Use Requests
Source: Equipment and Processing Report,
Click here