Understanding Biological Indicator Grow-Out Times
This study used biological indicators containing   Geobacillus stearothermophilus spores and a new technology to continuously monitor incubated BIs and record nonsterile results.


Pharmaceutical Technology
Volume 34, Issue 1

Biological indicators (BIs) are used globally to monitor the efficacy of sterilization processes for medical products (1–5). BIs contain high numbers—generally 104 to 106—of bacterial endo-spores that are highly resistant to the sterilization process for which they are designed. The spores are placed on or in a carrier material such as paper and packaged to protect the BI before as well as after sterilization and before culturing. BIs are useful for validation and routine monitoring of ethylene oxide, steam, dry heat, low temperature steam formaldehyde, vapor hydrogen peroxide, and related plasma sterilization systems.

In the United States, BIs sold for use in healthcare facilities are considered medical devices and are regulated by the US Food and Drug Administration good manufacturing practices (GMPs) as promulgated in the Quality System Regulation of 1996 (6). BIs sold to manufacturers of sterile medical products are considered manufacturing components. Regardless of the intended use, BIs must be manufactured under carefully controlled conditions to assure quality, reproducibility, and predictable performance (7).

Reducing biological indicator incubation time and the existing FDA protocol

How much incubation time is required to provide sufficient information to allow the user to make a correct decision regarding the acceptability of the sterilization process? In the mid-1980s, the FDA Center for Device and Radiological Health (CDRH) issued a protocol (hereafter termed the RIT [reduced incubation time] protocol) for BI manufacturers and users to follow to reduce the incubation time for BIs from the normal seven days to a shorter time based on the results of testing defined in the protocol (8). The RIT protocol requires manufacturers to expose 100 BIs from each of three lots to separate sterilization processes; the results of the exposure to the sterilization processes must result in 30 to 80 nonsterile BIs from each lot after incubation at the appropriate conditions for seven days. Achieving this outcome requires carefully-controlled reduced lethality (fractional) sterilization runs for which highly specialized sterilization equipment is often used.


Table I : Example of data necessary to meet the FDA CDRH criteria of determining a reduced incubation time of less than 7 days.
To determine a reduced incubation time for a BI/sterilization process according to the RIT protocol, the number of nonsterile BIs for each of the three lots is recorded after periodic reviews up to seven days of incubation; the number of nonsterile BIs at seven days is considered "100%." The number of nonsterile BIs at incubation times less than seven days is compared with the respective seven-day value, and a percentage is calculated; the qualified reduced incubation time is the longest time, for one or more of the three lots, to first exhibit a number of nonsterile BIs that is ≥ 97.0% of that observed after seven days of incubation. A typical example appears in Table I with the BIs being scored daily.

As required, for each of the three lots tested, the number of nonsterile BIs after seven days of incubation was greater or equal to 30 and less than or equal to 80. For lots 2 and 3, the 97.0% criterion was met after one day of incubation; for lot 1, however, the 97.0% criterion was met after two days of incubation. Based upon these results, a reduction of incubation time from seven to two days is acceptable.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters

Subscribe: Click to learn more about the newsletter
| Weekly
| Monthly
|Monthly
| Weekly

Survey
How does your company apply quality-by-design (QbD) principles to manufacturing processes?
To all processes for both new and legacy products
To all process for new products only
To select process for new products only
To select processes for both new and legacy products
Do not use QbD
To all processes for both new and legacy products
18%
To all process for new products only
13%
To select process for new products only
22%
To select processes for both new and legacy products
22%
Do not use QbD
24%
View Results
UPCOMING CONFERENCES

Programs for Investigational and Pre-Launch Drugs
Philadelphia, PA
July 17-18, 2013
Request Brochure

Strategic Pipeline Planning & Portfolio Valuation
Philadelphia, PA
August 13-14, 2013
Request Brochure

MES 2013 - Forum on Manufacturing Execution Systems
Philadelphia, PA
August 14-15, 2013
Request Brochure

Mobile Innovation for the Life Sciences Industry
Philadelphia, PA
August 20-21, 2013
Request Brochure

See All Conferences >>

Eric Langer Outsourcing Outlook Eric LangerOutsourcing's Modest Role as a Cost-Containment Strategy
Patricia Van Arnum Ingredients Insider Patricia Van ArnumIntellectual Property Battles in Solid-State Chemistry
Nathan Jessop Industry Insider Nathan Jessop Campaign Against Counterfeit Drugs Continues
Lynn Torbeck Statistical Solutions Lynn D. TorbeckCompositing Samples and the Risk to Product Quality
 More
Inadequate Access to Medicines Puts EU at Risk
FDA Offers Insight on QbD for Modified-Release Products
Global Biosimilars Market to Reach $2.445 Billion in 2013
Adapting to Change
AstraZeneca and Exco InTouch Collaborate to Augment Current COPD Pathways
FindPharma Custom Search
Source: Pharmaceutical Technology,
Click here