Erratic Recovery-Test Results - Pharmaceutical Technology

Latest Issue
PharmTech

Latest Issue
PharmTech Europe

Erratic Recovery-Test Results

Equipment and Processing Report

Untitled Document

Q: We’re attempting to validate a clean-in-place process using total organic carbon (TOC) as the analytical method. Our swab-sample recovery tests are giving us inconsistent results. Recovery values range from 95% to 205%, and there seems to be no correlation with the amount of drug product deposited on the coupon. What could be the problem?

 

A: TOC, unlike high-performance liquid chromatography (HPLC), is a nonspecific analytical method. The advantage in cleaning validation is that TOC can detect excipients, partial reactants, and detergents, in addition to the drug product. When coupled with visual inspection and other indirect tests, TOC can help minimize the overall contamination of the final product. However, because TOC is a nonspecific method, you need to account for and minimize potential sources of error to achieve consistent results. The erratic recovery in your studies is likely associated with one of these errors.

First, the measurement device should be capable of achieving the desired level of accuracy, sensitivity, specificity, and reproducibility per CFR 211.165. Guidance in US Pharmacopeia <1225> and the International Conference on Harmonization’s Q2 (R1) can help validate the analytical method. Use spike samples (i.e., pure water samples spiked with a known quantity of the target organic compound) to challenge the instrument accuracy before performing the swab-recovery tests. The swabs, vials, and coupons also can influence the consistency of results. Organic contaminants on these materials can contribute to the overall TOC result. Be sure to use swabs that have been certified for low TOC background. Swabs used in conjunction with HPLC methods often have high and variable organic content because it is not a crucial parameter for HPLC. Similarly, the vials selected for the testing should be certified for low TOC content. If you are cleaning the glassware yourself, use a mild caustic cleaning agent followed by thorough rinsing with low-TOC water. Finally, the coupons should be cleaned properly. They could be baked in a furnace to burn off residual organics or cleaned using a mild caustic detergent, then rinsed thoroughly with low-TOC water.

—Kevin Aumiller, senior applications specialist at GE Analytical Instruments

 

If you have a problem with your equipment or process, an industry expert may have the solution. Please send your question to Erik Greb, editor of Equipment and Processing Report, and we may be able to provide an answer in a future issue. All questions will remain anonymous.

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters

Subscribe: Click to learn more about the newsletter
| Weekly
| Monthly
|Monthly
| Weekly

Survey
FDASIA was signed into law two years ago. Where has the most progress been made in implementation?
Reducing drug shortages
Breakthrough designations
Protecting the supply chain
Expedited reviews of drug submissions
More stakeholder involvement
Reducing drug shortages
70%
Breakthrough designations
4%
Protecting the supply chain
17%
Expedited reviews of drug submissions
2%
More stakeholder involvement
7%
View Results
Eric Langerr Outsourcing Outlook Eric LangerRelationship-building at Top of Mind for Clients
Cynthia Challener, PhD Ingredients Insider Cynthia ChallenerRisk Reduction Top Driver for Biopharmaceutical Raw Material Development
Jill Wechsler Regulatory Watch Jill Wechsler Changes and Challenges for Generic Drugs
Faiz Kermaini Industry Insider Faiz KermainiNo Signs of a Slowdown in Mergers
From Generics to Supergenerics
CMOs and the Track-and-Trace Race: Are You Engaged Yet?
Ebola Outbreak Raises Ethical Issues
Better Comms Means a Fitter Future for Pharma, Part 2: Realizing the Benefits of Unified Communications
Better Comms Means a Fitter Future for Pharma, Part 1: Challenges and Changes
Source: Equipment and Processing Report,
Click here