Frequency Modulation Spectroscopy
A Novel Nondestructive Approach for Measuring Moisture Activity in Pharmaceutical Samples


Pharmaceutical Technology



Table V: Summary of key attributes of the various headspace-moisture techniques examined in the study.
Measurement precision and range. To evaluate measurement precision, 10 consecutive headspace moisture measurements of a 50-mL clear glass vial containing 16 capsules of a gelatin encapsulated dry product were taken at 40 C. Measurement precision, expressed by the percent relative standard deviation (% RSD) of the 10 consecutive readings, was 1.1%.

Measurement precision also was evaluated by repetitive measurements of several saturated salt solutions at 25 and 40 C. An individual vial containing a specified saturated salt solution was prepared for each measurement (see Tables VII and VIII). As Figures 6a and 6b show, % RSD values increased as the % RH value decreased.


Table VI: Examination of the specificity of the FMS 1400.
The ability of the FMS 1400 headspace-moisture analyzer to accurately measure relative humidity values close to 0% RH and 100% RH also was evaluated (see Table VI and Figure 6b). Five 50-mL clear vials containing USP water (with an expected relative humidity of 100%) and equilibrated at 40 C were measured. The results (mean = 98.8%, RSD = 0.4%) suggested that the Lighthouse instrument can measuring up to 100% RH with a high degree of accuracy. At the low end, five 50-mL clear vials containing dried desiccants (with an expected relative humidity of 0%) and equilibrated at 40 C were measured. The results (mean = 0.8%, RSD = 37.5%) suggested acceptable performance at the low end of the expected relative humidity range as well, although with less precision.

Conclusion


Table VII: Accuracy and linearity for measurements at 25 C for the FMS 1400.
This article introduced a novel nondestructive technique that is based on frequency modulation spectroscopy for the measurement of moisture activity in optically transparent packages. The technique appears to be a promising tool with reasonable method specificity, measurement precision, method precision, accuracy, linearity, range, and excellent analyst-to-analyst and time stability. The ability to make moisture activity measurements nondestructively for samples in a sealed glass container also enables new applications and offers significant advantages over traditional techniques.


Table VIII: Accuracy and linearity for measurements at 40 C for the FMS 1400.
The sealed system avoids all sample-handling issues during the measurement because the samples are not exposed to the environment. Furthermore, there is no interference from preceding measurements or concerns for leakage (because the sealed stopper-vial system provides an excellent barrier). Some applications enabled by this technique not offered by traditional techniques include measurement of moisture activity in optically transparent packages (e.g., lyophilized products) without breach of package, measurements of samples sealed in an inert gas atmosphere, and monitoring of moisture exchange between samples in real time. The high cost of the instrument will, however, restrict widespread application of this technique.

Acknowledgments


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters

Subscribe: Click to learn more about the newsletter
| Weekly
| Monthly
|Monthly
| Weekly

Survey
FDASIA was signed into law two years ago. Where has the most progress been made in implementation?
Reducing drug shortages
Breakthrough designations
Protecting the supply chain
Expedited reviews of drug submissions
More stakeholder involvement
Reducing drug shortages
27%
Breakthrough designations
9%
Protecting the supply chain
41%
Expedited reviews of drug submissions
9%
More stakeholder involvement
14%
View Results
Jim Miller Outsourcing Outlook Jim Miller Health Systems Raise the Bar on Reimbursing New Drugs
Cynthia Challener, PhD Ingredients Insider Cynthia ChallenerThe Mainstreaming of Continuous Flow API Synthesis
Jill Wechsler Regulatory Watch Jill Wechsler Industry Seeks Clearer Standards for Track and Trace
Siegfried Schmitt Ask the Expert Siegfried SchmittData Integrity
Sandoz Wins Biosimilar Filing Race
NIH Translational Research Partnership Yields Promising Therapy
Clusters set to benefit from improved funding climate but IP rights are even more critical
Supplier Audit Program Marks Progress
FDA, Drug Companies Struggle with Compassionate Use Requests
Source: Pharmaceutical Technology,
Click here