This article provides a historical review of computer validation in the pharmaceutical industry within the last three decades, evolving from the early years' initial concept and approach to today's current practices. Also included is how the regulations and industry have progressed in addressing the topic of computer validation.
We can understand FDA concerns about computer systems as we look at the following examples. The "Therac-25" medical-device
software went awry between 1983 and 1987, overdosing patients with X-rays (21). FDA stopped the use of the device. According
to the recollection of the authors, Wyeth Laboratories Inc. is believed to have received the first computer validation–related
483 from Center for Drug Evaluation and Research (CDER) in October 1983 for a lack of documentation for the validation of
a computer system used for the statistical analysis of data. In the late 1980s, the generic-drug scandal caused concern because
data had been falsified. One of the authors served as a witness in a grand jury investigation of the generic-drug scandal.
Interestingly, Phil Piasecki, a former FDA official who later worked in the industry, once told one of the authors of an observation
he made on computer systems. He noticed that the red light of a dehumidifier in a data center was on. After inquiring about
the dehumidifier's purpose, he cited the company for not having a standard operating procedure (SOP) for its use and maintenance.
These examples illustrate the wide range of enforcement actions and levels of impact on patients' safety (from serious to
limited and indirect impact). Since the mid 1990s, we have also seen numerous observations relating to computer-validation
issues, ranging from "the system is not validated" to "the SOP is not followed." Since the early 2000s, however, it seems
the number of observations has decreased. Between 2000 and 2005, the number of Warning Letters issued decreased by 50% (22),
perhaps resulting from a shift in FDA's focus after the 9/11 attack in 2001 to homeland security and food safety (23). At
the same time, FDA faces a decrease in budget spending (24).
FDA's presentation titled "Data Integrity, Another Looming Crisis," revealed its recent enforcement has focused on record
integrity (25). The presentation made it clear that based on recent inspection findings, FDA would refocus on the integrity
of e-records and train inspectors in this topic. Despite challenges that FDA faces, it has increased significantly the understanding
and knowledge of computer systems in the industry. The agency actively participates and has a Computer System National Expert
representing FDA in industry forums (e.g., GAMP), as well as providing computer validation training courses internally and
through selected providers. Hence, although we are now seeing fewer computer-validation 483s, partly because FDA is citing
predicate rules instead of Part 11, these 483s are more meaningful and should not be interpreted as a reflection of relaxed
enforcement.
Perspective on computer-validation practices
The Standish Group's "Chaos" reports, indicate that "incomplete requirements and specifications" and "changing requirements
and specifications" are two of the top three "project challenged factors" for computer-system projects (26). This assessment
seems to hold true in today's computer-validation practices. An informal poll conducted by the authors indicates that user
requirements are the main challenge when validating a computer system. User requirements have been a factor since the early
years of computer validation. Nonetheless, the general understanding about how to conduct computer validation in the industry
has increased since the 1990s, and most companies now have groups or departments dedicated to computer validation. Since the
introduction of the Sarbanes–Oxley financial regulations for a publicly traded company, more and more information technology
(IT) departments have established compliance groups. It is generally accepted now that IT infrastructure must be qualified
to meet growing regulatory requirements across the business.
Rory Budihandojo is a computer validation manager at Boehinger-Ingelheim, 2820 N. Normandy Dr., Petersburg, VA 23805, tel. 804.504.8373, fax 804.504.8637, rbudihandojo@bichemicals.com.
He is also a member of Pharmaceutical Technology's Editorial Advisory Board.
Articles by Rory Budihandojo
Steve Coates
Steve Coates is the director of computer system quality assurance at Wyeth.