Biological Indicator Growout Times: The Impact on Establishing a Reduced Incubation Time Protocol
The authors provide a review of test methodology and standards, including current industry and regulatory proposals, for biological indicator growout times.


Pharmaceutical Technology
Volume 36, Issue 5, pp. s42-s45

A biological indicator is defined in the International Organization for Standardization (ISO) 11139 terminology document as: "a test system containing viable microorganisms providing a defined resistance to a specified sterilization process" (1). Biological indicators (BIs) are used by medical-device manufacturers, pharmaceutical manufacturers, and healthcare institutions to provide measured response of the efficacy of sterilization processes; the sterility of medical items cannot be inspected or tested in a practical manner. BIs use known quantities of micro-organisms with a high degree of resistance to specific sterilization processes to generate quantitative information regarding efficacy of the processes.

Introduction

The presence of viable micro-organisms in or on a BI makes this form of sterilization indicator an appropriate tool to evaluate the sterilization process. BIs are unique in comparison with chemical indicators and physical monitors, which are also valuable and provide immediate or real-time information respectively. The latter monitors, however, do not directly provide a biological result.

That said, the use of viable micro-organisms requires a post-exposure incubation period to allow for growth of any test organisms that may have survived the sterilization process. This time delay in obtaining feedback from the sterilization cycle is often viewed as a major limitation to the use of BIs, particularly in applications for routine monitoring.

Viable micro-organisms, by nature, require time under suitable conditions of temperature, oxygen tension, nutrients, water, and so forth, to multiply and subsequently create a visual indication of their viability. The result of viability is typically demonstrated by turbidity in a broth culture medium, colonies on an agar surface, or a pH indicator or other metabolic-derived color changes that signify microbial metabolism and growth. A BI that is positive at any time during the predetermined incubation period that growth is detected is a straightforward result (i.e. a positive is a positive regardless of the incubation time). However, a negative BI result is ambiguous relative to the length of incubation. Because today's BIs use well-characterized bacterial endospores, all of the necessary conditions for germination and growth are established. Additionally, the performance requirements of BIs have been well characterized in the ISO 11138 BI Standards series (2). The only BI parameter that is not universally accepted to date is the length of the incubation time necessary to establish viability of BI organisms.

The release of medical items as "sterile" in both industry and in healthcare facilities is commonly based on negative BI monitoring results. The time from completion of the sterilization process to release of goods for distribution or patient use can have major economic and patient care implications. The incubation time at which a BI can be declared negative is therefore a significant issue and is to some extent complicated by the fact that any scientific conclusion based on a negative result is always open to debate.

The current ISO 14161:2009 standard, which provides guidance for users of BIs, recommends an incubation period of 7 days for established sterilization processes, such as ethylene oxide and moist heat, and 14 days for nonstandard or new sterilization processes (3). This guidance is consistent with the requirements of the BI manufacturer's standard ISO 11138-1:2006, but the 11138-1 document also specifies that the BI incubation time and temperature be validated (4). Additional text in ISO 14161:2009 regarding BI incubation time (12.3.3) states that:

  • A BI manufacturer may validate a reduced incubation time (RIT) for its spore carrier/recovery system (i.e., less than 7 days).
  • The RIT validation for this product does not have to be repeated by the end user as long as the BI is used with the same sterilization agent.
  • A spore carrier/recovery system that does not have an RIT validated by the BI manufacturer should be validated by the end user employing a defined statistical sampling plan and procedure with pre-established acceptance criteria (3).

Neither ISO BI document provides information on test methodology for validating BI incubation time. This lack of detail has led to confusing and nonuniform requirements for RIT documentation that have historically been applied to medical device and pharmaceutical manufacturers throughout the world.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters

Subscribe: Click to learn more about the newsletter
| Weekly
| Monthly
|Monthly
| Weekly

Survey
How does your company apply quality-by-design (QbD) principles to manufacturing processes?
To all processes for both new and legacy products
To all process for new products only
To select process for new products only
To select processes for both new and legacy products
Do not use QbD
To all processes for both new and legacy products
20%
To all process for new products only
13%
To select process for new products only
24%
To select processes for both new and legacy products
20%
Do not use QbD
22%
View Results
UPCOMING CONFERENCES

Programs for Investigational and Pre-Launch Drugs
Philadelphia, PA
July 17-18, 2013
Request Brochure

Strategic Pipeline Planning & Portfolio Valuation
Philadelphia, PA
August 13-14, 2013
Request Brochure

MES 2013 - Forum on Manufacturing Execution Systems
Philadelphia, PA
August 14-15, 2013
Request Brochure

Mobile Innovation for the Life Sciences Industry
Philadelphia, PA
August 20-21, 2013
Request Brochure

See All Conferences >>

Eric Langer Outsourcing Outlook Eric LangerOutsourcing's Modest Role as a Cost-Containment Strategy
Patricia Van Arnum Ingredients Insider Patricia Van ArnumIntellectual Property Battles in Solid-State Chemistry
Nathan Jessop Industry Insider Nathan Jessop Campaign Against Counterfeit Drugs Continues
Lynn Torbeck Statistical Solutions Lynn D. TorbeckCompositing Samples and the Risk to Product Quality
 More
Inadequate Access to Medicines Puts EU at Risk
FDA Offers Insight on QbD for Modified-Release Products
Global Biosimilars Market to Reach $2.445 Billion in 2013
Adapting to Change
AstraZeneca and Exco InTouch Collaborate to Augment Current COPD Pathways
FindPharma Custom Search
Source: Pharmaceutical Technology,
Click here